Metal Analyzer


ELECTROCHEMICAL TRACE METAL ANALYZER - IONIX



In compliance with the current regulations (i.e. EC REGULATION n° 466/2001) on the maximum concentration level, heavy metal detection in waters, air, ground, food, oil and cosmetic fields has become fundamental for the public and environmental health.

IONIX is a high-sensitivity analysis unit able to detect metal traces in different simple or complex matrixes. High speed to measure the potential variations, with their subsequent processing, allows to reach a high detectability level (often below ppb). Thus, very small metal traces can be accurately and repeatedly detected and quantified.

Unlike other similar techniques, analyses can be run on the ORIGINAL sample too, that is matrixes do not require any chemical or physical pre-treatment; moreover, samples are not destroyed during analyses. This makes IONIX particularly suitable for routine analyses and for use by non-particularly skilled operators. Spot on-site analyses can be performed thanks to its small compact size.

UPDATED TECHNIQUES:
  • PSA (Anodic Stripping Analysis)
  • CCSA (Constant current stripping analysis)
  • DPS (Differential pulse)
  • SWS (Square Wave)
  • CYU (Cyclic Voltammetry)

HEAVY METALS AND OTHER ORGANIC COMPOUNDS DETECTABLE:

Aluminum - Antimony - Arsenic - Bismuth - Cadmium - Cobalt - Chromium - Iron - Gallium - Manganese - Mercury - Molybdenum - Nickel - Lead - Copper - Selenium - Tin - Thallium - Titanium
Uranium - Vanadium - Zinc.

NOTE: Detectability limits and actual analyzability of the heavy metals and other organic compouns listed above depend on the kind of matrix and treatment conditions (e.g. mineralization, reagents purity, laboratory environmental conditions, operator's skills).

GENERAL SPECIFICATIONS
  • Power supply: 115 -230 Vac +/-10% 50-60 Hz 35VA
  • Dimensions WxLxH: 200 x 230 x 350 mm
  • Weight:8.5 kg
  • Minimum PC Requirements
  • Processor: Pentium® III
  • System memory: 1Gb RAM (52Gb recommended)
  • Hard disk free space: 100 Mb
  • CD-ROM unit: Yes
  • Serial port: N° 2 serial ports (RS-232 or USB)
  • Monitor: VGA monitor (1280x1024) or more
  • Printer: Any Windows" compatible printer
  • Operating system: Microsoft® Windows" XP
ELECTRICAL SPECIFICATIONS
  • Output voltage: ±12V
  • Response time: d100µs
  • Output impedance: e100M&
  • Current ranges: ±2mA up to ±2nA
  • Resolution: 16 bit
  • Input voltage range: ±10V
  • Input impedance: ±1012 &
  • Input leakage current: d10pA
  • Input noise: d25µVpp
  • Analog-digital conversion
  • Potential range: ±4096 mV
  • Resolution: 16 bit (±125 µV)
  • Conversion speed: e100 K sample/s
COMMUNICATION
Interface: Serial opto-isolated RS 232 C and opto-isolated USB

ELECTROCHEMICAL MATERIALS AND DEVICES
  • Electrodes support material: ARNITE®
  • Sample cell material: Borosilicate glass
  • Stirrers:Magnetic/with borosilicate glass helix Software controlled speed
  • Helix: Borosilicate glass
  • Purging system: Two-way automatic gas bubbler system
ELECTRODES
  • Type: 6 mm glass body with standard N6 conical joint
  • Reference Electrode: Ag - AgCl
  • Counter Electrode: Metal platinum (tip)
  • Working Electrode: 3mm Glassy carbon disk (GC V-10 grade)
  • Optional working electrodes: Gold tip, GC tip for RDE, Au tip for RDE, Hanging drop Hg (HDME)


 
 
 
 